Identificador persistente para citar o vincular este elemento:
http://hdl.handle.net/10553/51708
Título: | Dysfunction of an On-X Heart Valve by Pannus | Autores/as: | Abad, Cipriano Urso, Stefano Gomez, Elsa De la Vega, Maria |
Clasificación UNESCO: | 32 Ciencias médicas 321307 Cirugía del corazón |
Palabras clave: | Bioprosthesis Mitral valve Thrombosis Postoperative complications |
Fecha de publicación: | 2016 | Publicación seriada: | Journal of Heart Valve Disease | Resumen: | A 68-year-old woman with a history of previous double-valve replacement with On-X mechanical heart valves presented with clinical, echocardiographic and cardiac catheterization signs of obstruction of the On-X tricuspid heart valve prosthesis. The patient was successfully reoperated, but at surgery the valve was seen to be invaded by an abnormal overgrowth of pannus that blocked one of the leaflets. A small amount of non-obstructive fresh thrombus was also observed. The valve was successfully replaced with a biological heart valve prosthesis. The patient was discharged home, and is doing well four months after the operation, when echocardiography demonstrated normal function in the tricuspid valve. The present case represents the first ever report of pannus formation and subsequent dysfunction in an On-X heart valve, and also the first case of tricuspid valve malfunction and obstruction using this type of heart valve substitute. | URI: | http://hdl.handle.net/10553/51708 | ISSN: | 0966-8519 | Fuente: | The Journal of heart valve disease[ISSN 0966-8519],v. 25(5), p. 634-637 (Septiembre 2016) |
Colección: | Artículos |
Los elementos en ULPGC accedaCRIS están protegidos por derechos de autor con todos los derechos reservados, a menos que se indique lo contrario.