Identificador persistente para citar o vincular este elemento:
http://hdl.handle.net/10553/48061
Título: | A thermal model of a flow calorimeter | Autores/as: | Socorro, F. Rodríguez De Rivera, M. Jesus, Ch |
Clasificación UNESCO: | 2213 Termodinámica 220408 Líquidos |
Palabras clave: | Conduction calorimetry Flow calorimetry Liquid mixtures Modelling Signal processing, et al. |
Fecha de publicación: | 2001 | Publicación seriada: | Journal of Thermal Analysis and Calorimetry | Conferencia: | 12th ICTAC Congress 12th International Congress on Thermal Analysis and Calorimetry |
Resumen: | This work shows the difference between the results obtained in the electrical calibration and those corresponding to the chemical calibration in a TAM2277-204 flow-calorimeter by Thermometric. This difference is due to the fact that the mixture and the electrical dissipation do not occur in the same place. There are other additional aspects inherent to the mixture dissipation: the mixture does not take place instantaneously and extends in space. The physical model that has been calculated in this work explains the characteristics that are proper to a mixture dissipation and the effects of the injection in a qualitative and quantitative way. | URI: | http://hdl.handle.net/10553/48061 | ISSN: | 1418-2874 | DOI: | 10.1023/A:1011582306796 | Fuente: | Journal of Thermal Analysis and Calorimetry [ISSN 1418-2874], v. 64, p. 357-366 |
Colección: | Actas de congresos |
Los elementos en ULPGC accedaCRIS están protegidos por derechos de autor con todos los derechos reservados, a menos que se indique lo contrario.