Please use this identifier to cite or link to this item:
http://hdl.handle.net/10553/73915
Title: | Peripheral stent recovery after failed intracoronary delivery | Authors: | Pan, Manuel Medina, Alfonso Romero Moreno, Miguel Ángel Suárez de Lezo, José Hernández, Enrique Pavlovic, Djordje Melián, Francisco Marrero, José Cabrera, José A. |
UNESCO Clasification: | 320501 Cardiología | Keywords: | Coronary Palmaz-Schatz Stent Embolism Complications |
Issue Date: | 1992 | Journal: | Catheterization and Cardiovascular Diagnosis | Abstract: | The Palmaz-Schatz stent can be successfully deployed in most patients. However, in a small percentage of instances a systemic embolism of the stent has been reported. In the present article we describe an easy and fast method of stent capture when detachment from the balloon has happened (pulling back the stent-loaded balloon into the guiding catheter or femoral sheath). In this situation we propose the use of the coronary guide-wire "as a guide" to capture coaxially the lost stent. This method allows for continuing the procedure without removal of the femoral sheath. | URI: | http://hdl.handle.net/10553/73915 | ISSN: | 0098-6569 | DOI: | 10.1002/ccd.1810270316 | Source: | Catheterization and Cardiovascular Diagnosis [ISSN 0098-6569], v. 27 (3), p. 230-233, (Noviembre 1992) |
Appears in Collections: | Artículos |