Please use this identifier to cite or link to this item: http://hdl.handle.net/10553/73915
Title: Peripheral stent recovery after failed intracoronary delivery
Authors: Pan, Manuel
Medina, Alfonso
Romero Moreno, Miguel Ángel
Suárez de Lezo, José
Hernández, Enrique
Pavlovic, Djordje
Melián, Francisco
Marrero, José
Cabrera, José A.
UNESCO Clasification: 320501 Cardiología
Keywords: Coronary
Palmaz-Schatz Stent
Embolism
Complications
Issue Date: 1992
Journal: Catheterization and Cardiovascular Diagnosis 
Abstract: The Palmaz-Schatz stent can be successfully deployed in most patients. However, in a small percentage of instances a systemic embolism of the stent has been reported. In the present article we describe an easy and fast method of stent capture when detachment from the balloon has happened (pulling back the stent-loaded balloon into the guiding catheter or femoral sheath). In this situation we propose the use of the coronary guide-wire "as a guide" to capture coaxially the lost stent. This method allows for continuing the procedure without removal of the femoral sheath.
URI: http://hdl.handle.net/10553/73915
ISSN: 0098-6569
DOI: 10.1002/ccd.1810270316
Source: Catheterization and Cardiovascular Diagnosis [ISSN 0098-6569], v. 27 (3), p. 230-233, (Noviembre 1992)
Appears in Collections:Artículos
Show full item record

SCOPUSTM   
Citations

26
checked on Apr 14, 2024

WEB OF SCIENCETM
Citations

25
checked on Feb 25, 2024

Page view(s)

103
checked on Mar 9, 2024

Google ScholarTM

Check

Altmetric


Share



Export metadata



Items in accedaCRIS are protected by copyright, with all rights reserved, unless otherwise indicated.