Identificador persistente para citar o vincular este elemento: http://hdl.handle.net/10553/74272
Título: Near-infrared transmittance pulse oximetry with laser diodes
Autores/as: López Silva, Sonnia María 
Dotor Castilla, María Luísa
Silveira Martín, Juan Pedro
Clasificación UNESCO: 3311 tecnología de la instrumentación
Palabras clave: Oximetry
Sensors
Semiconductor lasers
Oxygen
Rhodamine B, et al.
Fecha de publicación: 2003
Publicación seriada: Journal of Biomedical Optics 
Resumen: Pulse oximeters are widely used for noninvasive monitoring of oxygen saturation in arterial blood hemoglobin. We present a transmittance pulse oximetry system based on near-infrared (NIR) laser diodes (750 and 850 nm) for monitoring oxygen saturation of arterial blood hemoglobin. The pulse oximetry system is made up of the optical sensor, sensor electronics, and processing block. Also, we show experimental results obtained during the development of the whole NIR transmittance pulse oximetry system along with modifications in the sensor configuration, signal processing algorithm, and calibration procedure. issues concerning wavelength selection and its implications for the improvement of the transmittance pulse oximetry technique are discussed. The results obtained demonstrate the proposed system's usefulness in monitoring a wide range of oxygen saturation levels.
URI: http://hdl.handle.net/10553/74272
ISSN: 1083-3668
DOI: 10.1117/1.1578495
Fuente: Journal of Biomedical Optics [ISSN 1083-3668], v. 8 (3), p. 525-533, (Julio 2003)
Colección:Artículos
miniatura
pdf
Adobe PDF (401,03 kB)
Vista completa

Citas SCOPUSTM   

30
actualizado el 17-nov-2024

Citas de WEB OF SCIENCETM
Citations

28
actualizado el 17-nov-2024

Visitas

89
actualizado el 04-may-2024

Descargas

315
actualizado el 04-may-2024

Google ScholarTM

Verifica

Altmetric


Comparte



Exporta metadatos



Los elementos en ULPGC accedaCRIS están protegidos por derechos de autor con todos los derechos reservados, a menos que se indique lo contrario.