Título: | Chronic Electrical Stimulation of the Otolith Organ: Preliminary Results in Humans with Bilateral Vestibulopathy and Sensorineural Hearing Loss | Autores/as: | Ramos Macias, Angel Ramos De Miguel, Ángel Rodriguez Montesdeoca,Isaura Borkoski Barreiro, Silvia Falcón González, Juan Carlos |
Clasificación UNESCO: | 32 Ciencias médicas | Palabras clave: | Balance Bilateral vestibulopathy Saccule Utricle Vestibular implant |
Fecha de publicación: | 2019 | Publicación seriada: | Audiology and Neuro-Otology | Resumen: | Introduction: Bilateral vestibulopathy is an important cause of imbalance that is misdiagnosed. The clinical management of patients with bilateral vestibular loss remains difficult as there is no clear evidence for an effective treatment. In this paper, we try to analyze the effect of chronic electrical stimulation and adaptation to electrical stimulation of the vestibular system in humans when stimulating the otolith organ with a constant pulse train to mitigate imbalance due to bilateral vestibular dysfunction (BVD). Methods: We included 2 patients in our study with BVD according to Criteria Consensus of the Classification Committee of the Bárány Society. Both cases were implanted by using a full-band straight electrode to stimulate the otoliths organs and simultaneously for the cochlear stimulation we use a perimodiolar electrode. Results: In both cases Vestibular and clinical test (video head impulse test, videonistagmography cervical vestibular evoked myogenic potentials, cVEMP and oVEMP), subjective visual vertical test, computerized dynamic posturography, dynamic gait index, Time UP and Go test and dizziness handicap index) were performed. Posture and gait metrics reveal important improvement if compare with preoperartive situation. Oscillopsia, unsteadiness, independence and quality of life improved to almost normal situation. Discussion/Conclusion: Prosthetic implantation of the otolith organ in humans is technically feasible. Electrical stimulation might have potential effects on balance and this is stable after 1 year follow-up. This research provides new possibilities for the development of vestibular implants to improve gravito-inertial acceleration sensation, in this case by the otoliths stimulation. | URI: | http://hdl.handle.net/10553/70142 | ISSN: | 1420-3030 | DOI: | 10.1159/000503600 | Fuente: | Audiology and Neurotology [ISSN 1420-3030] |
Colección: | Artículos |
Citas SCOPUSTM
31
actualizado el 17-nov-2024
Citas de WEB OF SCIENCETM
Citations
24
actualizado el 17-nov-2024
Visitas
97
actualizado el 02-mar-2024
Google ScholarTM
Verifica
Altmetric
Comparte
Exporta metadatos
Los elementos en ULPGC accedaCRIS están protegidos por derechos de autor con todos los derechos reservados, a menos que se indique lo contrario.