Identificador persistente para citar o vincular este elemento: http://hdl.handle.net/10553/48069
Título: A model-based study of deconvolution in time-varying or injection calorimeters
Autores/as: Marco, F.
De Rivera, M. Rodríguez 
Ortín, J.
Serra, T.
Torra, V.
Clasificación UNESCO: 2213 Termodinámica
Palabras clave: Calorimetry
Fecha de publicación: 1988
Editor/a: 0040-6031
Publicación seriada: Thermochimica Acta 
Resumen: A deconvolution algorithm based on time-varying inverse filtering is applied to a calorimetric model. The model is designed to simulate a device in which continuous liquid injection takes place. By means of a proper separation of different contributions to the thermogenesis, classical time-invariant deconvolution routines can be applied even if the model presents zeros in the time-invariant transfer function. The corrections achieved on the thermogenesis are evaluated for two different kinds of thermal power dissipation.
URI: http://hdl.handle.net/10553/48069
ISSN: 0040-6031
DOI: 10.1016/0040-6031(88)87262-9
Fuente: Thermochimica Acta [ISSN 0040-6031], v. 126, p. 171-179