Identificador persistente para citar o vincular este elemento:
http://hdl.handle.net/10553/48069
Título: | A model-based study of deconvolution in time-varying or injection calorimeters | Autores/as: | Marco, F. De Rivera, M. Rodríguez Ortín, J. Serra, T. Torra, V. |
Clasificación UNESCO: | 2213 Termodinámica | Palabras clave: | Calorimetry | Fecha de publicación: | 1988 | Editor/a: | 0040-6031 | Publicación seriada: | Thermochimica Acta | Resumen: | A deconvolution algorithm based on time-varying inverse filtering is applied to a calorimetric model. The model is designed to simulate a device in which continuous liquid injection takes place. By means of a proper separation of different contributions to the thermogenesis, classical time-invariant deconvolution routines can be applied even if the model presents zeros in the time-invariant transfer function. The corrections achieved on the thermogenesis are evaluated for two different kinds of thermal power dissipation. | URI: | http://hdl.handle.net/10553/48069 | ISSN: | 0040-6031 | DOI: | 10.1016/0040-6031(88)87262-9 | Fuente: | Thermochimica Acta [ISSN 0040-6031], v. 126, p. 171-179 |
Colección: | Artículos |
Citas SCOPUSTM
2
actualizado el 17-nov-2024
Citas de WEB OF SCIENCETM
Citations
2
actualizado el 17-nov-2024
Visitas
94
actualizado el 24-ago-2024
Google ScholarTM
Verifica
Altmetric
Comparte
Exporta metadatos
Los elementos en ULPGC accedaCRIS están protegidos por derechos de autor con todos los derechos reservados, a menos que se indique lo contrario.