Identificador persistente para citar o vincular este elemento:
http://hdl.handle.net/10553/46301
Título: | Development of a calorimetric sensor for medical application. Part II. Identification and simulation | Autores/as: | González De Chávez Menéndez, Jesús Socorro, F. Rodriguez de Rivera, M. |
Clasificación UNESCO: | 2210 Química física | Palabras clave: | Calorimetric sensor Calorimetric model Direct calorimetry Medical calorimetry Model |
Fecha de publicación: | 2013 | Publicación seriada: | Journal of Thermal Analysis and Calorimetry | Resumen: | A calorimetric sensor has been developed for local and direct measurement of calorific dissipations in different parts of the human body. The constructed prototype has a detection surface of 36 cm2. The calibration of the sensor is based on a semi-empirical model that permits to simulate the operation of the device, making easier an operational functioning method. The device is modeled as a system with two inputs and two outputs. The inputs are the calorific power (W) that is intended to be measured and the power (W pid) that dissipates a resistance, keeping constant the thermostat temperature through the use of a PID controller. The outputs are the thermostat temperature (T pid) and the calorimetric signal (y) that provides the thermopile that is in contact with the body. | URI: | http://hdl.handle.net/10553/46301 | ISSN: | 1388-6150 | DOI: | 10.1007/s10973-012-2701-z | Fuente: | Journal of Thermal Analysis and Calorimetry [ISSN 1388-6150], v. 113 (3), p. 1003-1007, (Septiembre 2013) |
Colección: | Artículos |
Citas SCOPUSTM
10
actualizado el 17-nov-2024
Citas de WEB OF SCIENCETM
Citations
8
actualizado el 17-nov-2024
Visitas
96
actualizado el 02-nov-2024
Google ScholarTM
Verifica
Altmetric
Comparte
Exporta metadatos
Los elementos en ULPGC accedaCRIS están protegidos por derechos de autor con todos los derechos reservados, a menos que se indique lo contrario.