dc.description.abstract | OBJECTIVE: This study aims to evaluate the increase in power of cochlear implants (CI) as the reference electrode (RE) position changes. Patients in whom it is necessary to use the full power of the device to achieve the desired stimulation levels, this strategy will ensure that stimulation capability is maximized.MATERIALS AND METHODS: The variability in RE placement in the temporal bone has a measurable effect on the electrical current, impedance, and power consumption, and if the electrode position has a functional effect on the stimulation intensity. The following three approaches were used: 1) classical circuit analysis, 2) 2-dimensional numerical simulations, and 3) real temporal bone measurements using a purpose-made CI.RESULTS: The three approaches demonstrate a significant decrease in the current intensity and electrical resistance for distances that are closer to the intra-cochlear electrode. The results also demonstrate that to maintain a constant current, shorter distances require 33% less power.CONCLUSION: Reference electrode position during surgery can make significant differences in CI power consumption and threshold intensity, which allows a more powerful stimulation in complicated patients (i.e., those with otosclerosis). This study presents an attractive perspective to surgeons, as it shows a way to decrease consumption that might result in a longer battery life or more power to be devoted to coding strategy performance. | en_US |
dc.language | eng | en_US |
dc.relation.ispartof | Journal of International Advanced Otology | en_US |
dc.source | Journal Of International Advanced Otology [ISSN 1308-7649], v. 11 (3), p. 222-228, (Diciembre 2015) | en_US |
dc.subject | 32 Ciencias médicas | en_US |
dc.subject.other | Design | en_US |
dc.subject.other | Configuration | en_US |
dc.subject.other | Stimulation | en_US |
dc.subject.other | Array | en_US |
dc.subject.other | Cochlear Implat Impedance | en_US |
dc.subject.other | Cochlear Implant | en_US |
dc.subject.other | Power Consumption | en_US |
dc.subject.other | Refernce Electrode | en_US |
dc.subject.other | Threshold Intensity | en_US |
dc.title | The effect of reference electrode position in cochlear implants | en_US |
dc.type | info:eu-repo/semantics/Article | en_US |
dc.type | Article | en_US |
dc.identifier.doi | 10.5152/iao.2015.1746 | en_US |
dc.identifier.scopus | 84975701641 | - |
dc.identifier.isi | 000371282300010 | - |
dc.contributor.authorscopusid | 56866155200 | - |
dc.contributor.authorscopusid | 6701550535 | - |
dc.contributor.authorscopusid | 57189888621 | - |
dc.contributor.authorscopusid | 56573990500 | - |
dc.identifier.eissn | 2148-3817 | - |
dc.description.lastpage | 228 | en_US |
dc.identifier.issue | 3 | - |
dc.description.firstpage | 222 | en_US |
dc.relation.volume | 11 | en_US |
dc.investigacion | Ciencias de la Salud | en_US |
dc.type2 | Artículo | en_US |
dc.contributor.daisngid | 11181863 | - |
dc.contributor.daisngid | 1190904 | - |
dc.contributor.daisngid | 11928510 | - |
dc.contributor.daisngid | 5221004 | - |
dc.description.numberofpages | 7 | en_US |
dc.utils.revision | Sí | en_US |
dc.contributor.wosstandard | WOS:Ramos-Miguel, A | - |
dc.contributor.wosstandard | WOS:Ramos-Macias, A | - |
dc.contributor.wosstandard | WOS:Artiles, JV | - |
dc.contributor.wosstandard | WOS:Zaballos, MTP | - |
dc.date.coverdate | Enero 2015 | en_US |
dc.identifier.ulpgc | Sí | es |
dc.description.sjr | 0,146 | |
dc.description.jcr | 0,123 | |
dc.description.sjrq | Q3 | |
dc.description.jcrq | Q4 | |
dc.description.scie | SCIE | |
item.fulltext | Sin texto completo | - |
item.grantfulltext | none | - |
crisitem.author.dept | GIR SIANI: Ingeniería biomédica aplicada a estimulación neural y sensorial | - |
crisitem.author.dept | IU Sistemas Inteligentes y Aplicaciones Numéricas | - |
crisitem.author.dept | Departamento de Ciencias Médicas y Quirúrgicas | - |
crisitem.author.orcid | 0000-0002-4709-5559 | - |
crisitem.author.parentorg | IU Sistemas Inteligentes y Aplicaciones Numéricas | - |
crisitem.author.fullName | Ramos Macías, Ángel Manuel | - |
Appears in Collections: | Artículos |
SCOPUSTM
Citations
5
checked on Mar 2, 2025
WEB OF SCIENCETM
Citations
5
checked on Mar 2, 2025
Page view(s)
127
checked on Jan 11, 2025
Google ScholarTM
Check
Altmetric
Share
Export metadata
Items in accedaCRIS are protected by copyright, with all rights reserved, unless otherwise indicated.