Title: | A model-based study of deconvolution in time-varying or injection calorimeters | Authors: | Marco, F. De Rivera, M. Rodríguez Ortín, J. Serra, T. Torra, V. |
UNESCO Clasification: | 2213 Termodinámica | Keywords: | Calorimetry | Issue Date: | 1988 | Publisher: | 0040-6031 | Journal: | Thermochimica Acta | Abstract: | A deconvolution algorithm based on time-varying inverse filtering is applied to a calorimetric model. The model is designed to simulate a device in which continuous liquid injection takes place. By means of a proper separation of different contributions to the thermogenesis, classical time-invariant deconvolution routines can be applied even if the model presents zeros in the time-invariant transfer function. The corrections achieved on the thermogenesis are evaluated for two different kinds of thermal power dissipation. | URI: | http://hdl.handle.net/10553/48069 | ISSN: | 0040-6031 | DOI: | 10.1016/0040-6031(88)87262-9 | Source: | Thermochimica Acta [ISSN 0040-6031], v. 126, p. 171-179 |