Identificador persistente para citar o vincular este elemento: http://hdl.handle.net/10553/45965
Título: Antiseptic chamber-containing hub reduces central venous catheter-related infection: A prospective, randomized study
Autores/as: León, Cristóbal
Álvarez-Lerma, Francisco
Ruiz-Santana, Sergio 
González, Víctor
De La Torre, María Victoria
Sierra, Rafael
León, Miguel
Rodrigo, Juan José
Clasificación UNESCO: 32 Ciencias médicas
3201 Ciencias clínicas
Palabras clave: Antiseptic chamber
Central venous catheter
Infection
Fecha de publicación: 2003
Publicación seriada: Critical Care Medicine 
Resumen: Objective: The hub connecting the catheter and the infusion equipment is a common portal of entry for bacteria causing catheter-related sepsis. We assessed the efficacy of a new hub model (Segur-Lock) that incorporates an antiseptic chamber filled with 3% iodinated alcohol in preventing endoluminal catheter contamination and catheter-related bloodstream infection arising at the hub. Design: Prospective, randomized, multicenter study. Setting: Seven medical and surgical adult intensive care units in Spain. Patients: A total of 230 patients aged 18 yrs or older requiring the insertion of a nontunneled central venous catheter for >or=6 days from January 1, 1998, to April 30, 1999. Interventions: Patients were randomized at the time of catheter insertion to receive catheters with standard Luer-lock connector (control group, n = 114) or catheters with the new hub model (n = 116). Measurements and main results: Skin, catheter tip, and hub cultures were performed at the time the catheter was withdrawn because therapy was terminated or due to suspicion of sepsis, in which case peripheral blood and infusate cultures were simultaneously taken. Catheter-related bloodstream infection was diagnosed in 19 (8.3%) patients. Catheters were more often withdrawn because of suspicion of infection in the control group (43.8%) than in the new hub model group (30.1%, p <.035). The prevalence of culture-positive catheter hubs without associated bacteremia (colonization) was higher in the control group (14.4% vs. 4.3%, p <.001). Catheter-related bloodstream infection from the catheter hub also occurred more frequently in controls than in patients assigned to the new hub model (7% vs. 1.7%; p <.049; relative risk, 4.14; 95% confidence interval, 0.8-19). Conclusions: This new antiseptic chamber-containing hub has proved to be effective in preventing endoluminal bacterial colonization and catheter-related bloodstream infection from hub contamination in intensive care unit patients with central venous catheters inserted for >or=6 days.
URI: http://hdl.handle.net/10553/45965
ISSN: 0090-3493
DOI: 10.1097/01.CCM.0000026327.58305.22
Fuente: Critical Care Medicine [ISSN 0090-3493], v. 31, p. 1318-1324
Colección:Artículos
Vista completa

Citas SCOPUSTM   

39
actualizado el 22-dic-2024

Citas de WEB OF SCIENCETM
Citations

36
actualizado el 22-dic-2024

Visitas

89
actualizado el 27-jul-2024

Google ScholarTM

Verifica

Altmetric


Comparte



Exporta metadatos



Los elementos en ULPGC accedaCRIS están protegidos por derechos de autor con todos los derechos reservados, a menos que se indique lo contrario.