Identificador persistente para citar o vincular este elemento:
http://hdl.handle.net/10553/42889
Title: | A new device for the fixation of anterior cruciate ligament tendon grafts. Design and experimental study | Authors: | Martel, O. Carta, J. A. Garcés, Gerardo L. |
UNESCO Clasification: | 320714 Osteopatología 331402 Prótesis |
Keywords: | Anterior cruciate ligament reconstruction Interference screw Implant design Mechanical testing Mechanical testing |
Issue Date: | 2007 | Publisher: | 1350-4533 | Journal: | Medical Engineering and Physics | Abstract: | In this paper the design and experimental analysis is presented of a new fixation device of anterior cruciate ligament (ACL) grafts of the knee. This device is inserted into the bone tunnel, after the graft, in the same way as an interference screw. However, the fixation device described in this paper has been designed in such a way that, after the insertion of a threaded element in its interior, some of its components expand in a radial direction, pressing against the walls of the bone tunnel and thereby increasing the fixation of the graft. This expansion device can be used in both the femur and the tibia. The device proposed in this paper was compared with an interference screw for load failure and fixation stiffness in experiments performed using porcine bones. The failure load was significantly higher in the new expansion device group (633 ± 202 N) than in the interference screw group (471 ± 179 N). The stiffness obtained when the new device was used (59 ± 20 N/mm) was also significantly higher than that obtained using the interference screw (37 ± 19 N/mm) (t-test, P < 0.05). According to these results, this new device could be considered a good alternative to improve fixation of anterior cruciate ligament grafts. | URI: | http://hdl.handle.net/10553/42889 | ISSN: | 1350-4533 | DOI: | 10.1016/j.medengphy.2006.01.013 | Source: | Medical Engineering and Physics [ISSN 1350-4533], v. 29, p. 163-168 |
Appears in Collections: | Artículos |
SCOPUSTM
Citations
7
checked on Mar 30, 2025
WEB OF SCIENCETM
Citations
7
checked on Mar 30, 2025
Page view(s)
111
checked on Jan 11, 2025
Download(s)
5
checked on Jan 11, 2025
Google ScholarTM
Check
Altmetric
Share
Export metadata
Items in accedaCRIS are protected by copyright, with all rights reserved, unless otherwise indicated.