Identificador persistente para citar o vincular este elemento: http://hdl.handle.net/10553/119404
Título: Concordance among measurements obtained by three pulse oximeters currently used by health professionals.
Autores/as: Díaz González, Candelaria Merced 
De La Rosa Hormiga, Milagros 
Ramal López, Josefa María 
Deniz Rivero, Yasmina
Marrero Morales, María Sandra 
Clasificación UNESCO: 32 Ciencias médicas
3201 Ciencias clínicas
Palabras clave: Gases
Monitoring
Oximetry
Pulse oximeter
Fecha de publicación: 2014
Publicación seriada: Journal of Clinical and Diagnostic Research 
Resumen: Introduction: Oxygen saturation is considered as the 5(th) vital sign. Presently, there exist fixed and wireless pulse oximeters, being the latter most widely used in the last years. Some of them have no possibility of calibration. This situation leads the health staff to adopt therapeutic attitudes which can be wrong. Therefore, it is extremely important to know if these wireless oximeters show a right concordance as regards measurements, since it is of great interest in daily clinical practice. Objective: To evaluate concordance among measurements obtained by three different pulse oximeters currently used by health professionals. Materials and methods: This is an observational, descriptive and cross-sectional study related to the concordance of the results obtained in measurements collected by three different pulse oximeters (one monitor and two wireless oximeters) which are available and in use in this hospital unit. The sample size calculation was performed for a concordance above 0.81 and an estimation error which did not exceed 0.20. The intraclass correlation index (ICI) was used to establish the concordance whereas the Landis-Koch criteria were used to interpret the results. Systematic errors were analyzed using the Bland-Altman plot. Results: The overall concordance among the three pulse oximeters analyzed resulted in 0.88, a value considered as "good" according to the Landis-Koch criteria. Conclusion: The results obtained show that in daily clinical practice both wireless pulse oximeters analyzed can be used with a certain reliability, taking into account the limitations of this research.
URI: http://hdl.handle.net/10553/119404
ISSN: 0973-709X
DOI: 10.7860/JCDR/2014/8874.4763
Fuente: Journal of Clinical and Diagnostic Research [0973-709X], v. 8(8), pp.9-12 (Agosto 2014)
Colección:Artículos
Adobe PDF (256,03 kB)
Vista completa

Citas de WEB OF SCIENCETM
Citations

4
actualizado el 15-dic-2024

Visitas

30
actualizado el 06-abr-2024

Descargas

7
actualizado el 06-abr-2024

Google ScholarTM

Verifica

Altmetric


Comparte



Exporta metadatos



Los elementos en ULPGC accedaCRIS están protegidos por derechos de autor con todos los derechos reservados, a menos que se indique lo contrario.