Title: | Evolution and testing of a metallic ebulliometer for VLE determination | Authors: | Susial Badajoz, Pedro García Montesdeoca, María Isabel García Vera, Diego Marrero Pérez, Anibal Santiago Garcia, Dunia Esther |
UNESCO Clasification: | 2213 Termodinámica | Keywords: | 1-Propanol 2-Propanol Isobutyl Acetate Pc-Saft Vapor-Liquid Equilibri |
Issue Date: | 2022 | Journal: | Iranian Journal of Chemistry and Chemical Engineering | Abstract: | An ebulliometer designed with automated feeding was tested and verified in this work. The ebulliometer is made of stainless steel. The feeding of the substances is automatically controlled with a computer. In the equipment both the liquid and vapor phases are recirculated. The binary mixtures isobutyl acetate + 1–propanol and isobutyl acetate + 2–propanol were studied and the vapor–liquid equilibrium of these mixtures was determined at 101 kPa. The isobaric T–x–y data are reported, including the azeotropic point of the isobutyl acetate + 1–propanol binary system. The calculations have been carried out considering the non–ideal vapor phase and the activity coefficients of the liquid phase have been obtained. The thermodynamic consistency of the system was verified with the Van Ness point–to–point test. An ebulliometer designed with automated feeding was tested and verified in this work. The ebulliometer is made of stainless steel. The feeding of the substances is automatically controlled with a computer.(PC–SAFT). | ISSN: | 1021-9986 | DOI: | 10.30492/IJCCE.2022.556040.5393 | Source: | Iranian Journal of Chemistry and Chemical Engineering [ISSN 1021-9986], v. 42, nº 6, p. 1925-1937, (Junio 2023) |
Appears in Collections: | Artículos |
Page view(s)
80
checked on Oct 5, 2024
Download(s)
23
checked on Oct 5, 2024
Google ScholarTM
Check
Altmetric
Share
Export metadata
Items in accedaCRIS are protected by copyright, with all rights reserved, unless otherwise indicated.