Identificador persistente para citar o vincular este elemento:
http://hdl.handle.net/10553/48049
Título: | Signal processing and uncertainty in an isothermal titration calorimeter | Autores/as: | Rodríguez De Rivera, M. Socorro, F. |
Clasificación UNESCO: | 2213 Termodinámica | Palabras clave: | Calibration Isothermal titration calorimetry Liquid mixtures Uncertainty |
Fecha de publicación: | 2007 | Editor/a: | 1388-6150 | Publicación seriada: | Journal of Thermal Analysis and Calorimetry | Conferencia: | 9th European Symposium on Thermal Analysis and Calorimetry | Resumen: | In this paper, it is made a study of the accuracy of an isothermal titration calorimeter in the operating mode of ‘continuous injection’. The experimental equipment has been a TAM2277-201/2250 by Thermometric AB and the liquid mixtures used in the calibration have been the mixture cyclohexane+benzene and the mixture water+ethanol. The calibration contemplates different effects that affect the uncertainty in the determination of the sensitivity, the effect of the liquid injection, the treatment of the calorimetric signal, the variation of the experimental baseline and the different noises included in the calorimetric signal. | URI: | http://hdl.handle.net/10553/48049 | ISSN: | 1388-6150 | DOI: | 10.1007/s10973-006-8104-2 | Fuente: | Journal of Thermal Analysis and Calorimetry [ISSN 1388-6150], v. 88, p. 745-750 |
Colección: | Artículos |
Los elementos en ULPGC accedaCRIS están protegidos por derechos de autor con todos los derechos reservados, a menos que se indique lo contrario.