Please use this identifier to cite or link to this item:
http://hdl.handle.net/10553/41457
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Díaz González, Candelaria Merced | en_US |
dc.contributor.author | de la Rosa-Hormiga, Milagros | en_US |
dc.contributor.author | Ramal López, Josefa María | en_US |
dc.contributor.author | González-Henríquez, Juan José | en_US |
dc.contributor.author | Marrero Morales, Sandra | en_US |
dc.date.accessioned | 2018-07-04T07:07:47Z | - |
dc.date.available | 2018-07-04T07:07:47Z | - |
dc.date.issued | 2018 | en_US |
dc.identifier.issn | 1365-2702 | en_US |
dc.identifier.uri | http://hdl.handle.net/10553/41457 | - |
dc.description.abstract | Aims and objectives: To evaluate both concordance among those measurements obtained by three different pulse oximeters currently used by nursing professionals in hospital units and the factors which can influence this concordance. Background: Many models of wireless pulse oximeters in the present market do not offer possibility of calibration and, therefore, they do not ensure patients′ safety in daily clinical practice. Design: This is a descriptive and cross-sectional study. The sample of patients (140) was selected from such hospital units, and all of them had to fulfil inclusion and exclusion criteria for participation. Methods: The instruments used to carry out this research were a monitor (calibrated), two models of portable wireless pulse oximeters (used for 3 years by the nursing staff, without being calibrated), a tympanic thermometer and a weather meter. The concordance correlation coefficient (CCC) was used to establish the concordance, whereas the Landis-Koch criteria were used to interpret the results. Results: The concordance among the three devices was considered as “good” (CCC: 0.925 and 0.974 (95%)). The CCC (0.925) for saturation measures was regarded as “very good”/”almost perfect,” and the pulse measure was considered as “very good” CCC 0.974, providing in both cases a high level of concordance (CCC > 0.91). Conclusions: The overall concordance as regards pulse and oxygen saturation among the three pulse oximeters analysed is considered as “very good” according to the Landis-Koch criteria. Relevance to clinical practice: There exists a “very good” concordance among two wireless oximeters which have been used by the health staff for 3 years and which did not offer any possibility of calibration and one monitor which belongs to the healthcare institution. This aspect is of crucial importance in daily clinical practice, and it is also relevant to ensure patient's safety. | en_US |
dc.language | eng | en_US |
dc.relation.ispartof | Journal of Clinical Nursing | en_US |
dc.source | Journal of Clinical Nursing [ISSN 1365-2702], v. 27, p. 677-683 | en_US |
dc.subject | 32 Ciencias médicas | en_US |
dc.subject.other | Gases | en_US |
dc.subject.other | Monitoring | en_US |
dc.subject.other | Pulse oximetry | en_US |
dc.subject.other | Safety | en_US |
dc.subject.other | Transcutaneous | en_US |
dc.title | Factors which influence concordance among measurements obtained by different pulse oximeters currently used in some clinical situations | en_US |
dc.type | info:eu-repo/semantics/Article | en_US |
dc.type | Article | en_US |
dc.identifier.doi | 10.1111/jocn.14012 | en_US |
dc.identifier.scopus | 85042331580 | - |
dc.identifier.isi | 000425733600045 | - |
dc.contributor.authorscopusid | 57200720824 | - |
dc.contributor.authorscopusid | 56368199200 | - |
dc.contributor.authorscopusid | 57200727800 | - |
dc.contributor.authorscopusid | 6508312288 | - |
dc.contributor.authorscopusid | 57200721567 | - |
dc.description.lastpage | 683 | en_US |
dc.identifier.issue | 3-4 | - |
dc.description.firstpage | 677 | en_US |
dc.relation.volume | 27 | en_US |
dc.investigacion | Ciencias de la Salud | en_US |
dc.type2 | Artículo | en_US |
dc.contributor.daisngid | 2486836 | - |
dc.contributor.daisngid | 18007301 | - |
dc.contributor.daisngid | 21543554 | - |
dc.contributor.daisngid | 2563867 | - |
dc.contributor.daisngid | 13848742 | - |
dc.utils.revision | Sí | en_US |
dc.contributor.wosstandard | WOS:Diaz-Gonzalez, CD | - |
dc.contributor.wosstandard | WOS:de la Rosa-Hormiga, M | - |
dc.contributor.wosstandard | WOS:Ramal-Lopez, JM | - |
dc.contributor.wosstandard | WOS:Gonzalez-Henriquez, JJ | - |
dc.contributor.wosstandard | WOS:Marrero-Morales, MS | - |
dc.date.coverdate | Febrero 2018 | en_US |
dc.identifier.ulpgc | Sí | en_US |
dc.description.sjr | 0,768 | |
dc.description.jcr | 1,757 | |
dc.description.sjrq | Q1 | |
dc.description.jcrq | Q1 | |
dc.description.scie | SCIE | |
dc.description.ssci | SSCI | |
item.grantfulltext | open | - |
item.fulltext | Con texto completo | - |
crisitem.author.dept | GIR IUIBS: Nutrición | - |
crisitem.author.dept | IU de Investigaciones Biomédicas y Sanitarias | - |
crisitem.author.dept | Departamento de Enfermería | - |
crisitem.author.dept | GIR IUIBS: Nutrición | - |
crisitem.author.dept | IU de Investigaciones Biomédicas y Sanitarias | - |
crisitem.author.dept | Departamento de Enfermería | - |
crisitem.author.dept | GIR IUIBS: Nutrición | - |
crisitem.author.dept | IU de Investigaciones Biomédicas y Sanitarias | - |
crisitem.author.dept | Departamento de Enfermería | - |
crisitem.author.dept | GIR IUIBS: Rendimiento humano, ejercicio físico y salud | - |
crisitem.author.dept | IU de Investigaciones Biomédicas y Sanitarias | - |
crisitem.author.dept | Departamento de Matemáticas | - |
crisitem.author.dept | GIR Lingüística aplicada a la docencia de la lengua extranjera, su literatura y traducción | - |
crisitem.author.dept | Departamento de Filología Moderna, Traducción e Interpretación | - |
crisitem.author.orcid | 0000-0003-1797-4752 | - |
crisitem.author.orcid | 0000-0002-6768-0982 | - |
crisitem.author.orcid | 0000-0002-6409-1787 | - |
crisitem.author.orcid | 0000-0002-0622-5580 | - |
crisitem.author.orcid | 0000-0002-9120-3772 | - |
crisitem.author.parentorg | IU de Investigaciones Biomédicas y Sanitarias | - |
crisitem.author.parentorg | IU de Investigaciones Biomédicas y Sanitarias | - |
crisitem.author.parentorg | IU de Investigaciones Biomédicas y Sanitarias | - |
crisitem.author.parentorg | IU de Investigaciones Biomédicas y Sanitarias | - |
crisitem.author.parentorg | Departamento de Filología Moderna, Traducción e Interpretación | - |
crisitem.author.fullName | Díaz González, Candelaria Merced | - |
crisitem.author.fullName | De La Rosa Hormiga, Milagros | - |
crisitem.author.fullName | Ramal López, Josefa María | - |
crisitem.author.fullName | González Henríquez, Juan José | - |
crisitem.author.fullName | Marrero Morales, María Sandra | - |
Appears in Collections: | Artículos |
SCOPUSTM
Citations
2
checked on Dec 1, 2024
WEB OF SCIENCETM
Citations
1
checked on Nov 24, 2024
Page view(s)
204
checked on Jul 27, 2024
Download(s)
250
checked on Jul 27, 2024
Google ScholarTM
Check
Altmetric
Share
Export metadata
Items in accedaCRIS are protected by copyright, with all rights reserved, unless otherwise indicated.