Título: Can an expansion device be used in anterior cruciate ligament reconstruction? An in vitro study of soft tissue graft tibial fixation
Autores/as: Martel, Oscar 
Garcés Martín, Gerardo 
Yánez, Alejandro 
Cuadrado, Alberto 
Cárdenes, Juan F. 
Clasificación UNESCO: 3313 Tecnología e ingeniería mecánicas
Palabras clave: ACL reconstruction
Anterior cruciate ligament (ACL)
Biomechanical testing
Expansion device
Interference screw
Fecha de publicación: 2016
Publicación seriada: Knee 
Resumen: Background The purpose of this study was to compare the mechanical properties of an interference screw with an expansion device in anterior cruciate ligament (ACL) reconstruction. Methods A total of 52 porcine tibia and 20 polyurethane foam blocks (0.16 g/cm3) were used. Forty pullout tests were carried out to combine the two types of bones – surrogate and porcine – with the two fixation systems: interference screw and expansion device (n = 10 per group). Thirty-two cyclic tests (n = 8 per group) were carried out with both fixation devices in porcine bone at two different force amplitudes (100 N and 200 N). Results Stiffness and load values (mean ± SD) at six millimeters of displacement for the expansion device and the interference screw were 74 ± 33 N/mm, 318 ± 135 N, and 52 ± 28 N/mm, 205 ± 70 N, respectively, showing a difference in stiffness (P = 0.016) and load at six millimeters of displacement (P = 0.001). No correlation between insertion torque and the ultimate failure load was found for both fixation devices tested. In cyclic tests, significantly higher (P < 0.001) numbers of cycles (mean ± SD) were reached with the expansion device (81,014 ± 30,291 at 100 N; 13,462 ± 11,351 at 200 N) than with the interference screw (15,100 ± 8623 at 100 N; 343 ± 113 at 200 N) at six millimeters of displacement. Conclusion The use of an expansion device for ACL reconstructions seemed to be a promising alternative to an interference screw. Insertion torque alone was not a useful predictor of graft fixation strength in ACL reconstructions.
URI: https://accedacris.ulpgc.es/handle/10553/41371
ISSN: 0968-0160
DOI: 10.1016/j.knee.2016.06.009
Fuente: Knee[ISSN 0968-0160], v. 23(6), p. 1049-1054