Please use this identifier to cite or link to this item:
http://hdl.handle.net/10553/306
Title: | Signal processing and uncertainty in an isothermal titration | Authors: | Rodríguez de Rivera Rodríguez, Manuel Socorro Lorenzo, Fabiola |
UNESCO Clasification: | 2213 Termodinámica | Keywords: | Calibration Liquid mixtures Uncertainty Isothermal titration calorimetry |
Issue Date: | 2007 | Journal: | Journal of Thermal Analysis and Calorimetry | Abstract: | In this paper, it is made a study of the accuracy of an isothermal titration calorimeter in the operating mode of ‘continuous injection’. The experimental equipment has been a TAM2277-201/2250 by Thermometric AB and the liquid mixtures used in the calibration have been the mixture cyclohexane+benzene and the mixture water+ethanol. The calibration contemplates different effects that affect the uncertainty in the determination of the sensitivity, the effect of the liquid injection, the treatment of the calorimetric signal, the variation of the experimental baseline and the different noises included in the calorimetric signal. | URI: | http://hdl.handle.net/10553/306 | ISSN: | 1388-6150 | Source: | Journal of thermal analysis and calorimetry [ISSN 1388-6150], v. 88 (3), p. 745-750 |
Appears in Collections: | Artículos |